Lab material and scientific instruments.

At Glasatelier Saillart you will find a complete range of brand products at an attractive price!

TN500 White Light Turbidimeter with GLP Data Management and USB Output

The picture may be different from the article and can include supply, which is not in the standard scope.

The Apera Instruments TN500 provides fast and reliable measurements of the turbidity (turbidity) of liquids. The instrument is equipped with a tungsten filament lamp and operates via a 90° nephelometric scattering method that is consistent with U.S. EPA 180.1 method for determining turbidity in drinking, ground, surface, and saline waters, as well as in domestic and industrial wastes. It is well suited for rapidly settling samples or fluctuating or variable measurements thanks to TruRead™ measurement mode. The TN500 has a memory function for up to 200 data sets and can be connected to a PC via USB to read out the data or perform real-time storage. The large LCD with progress indicator and intuitive operating instructions in picture and text form make it user-friendly and easy to operate. It is suitable for use on site or in the laboratory.

Technical Specifications
Measuring method Nephelometric method (90°), U.S. EPA180.1 compliant
Light source Tungsten light bulb
Measuring range

0 ~ 1000 NTU (FNU), automatic range selection:

0.01 ~ 19.99 NTU (FNU)

20.0 ~ 99.9 NTU (FNU)

100 ~ 1000 NTU (FNU)

Accuracy ≤ ±2% of reading ± stray light
Reproducibility ≤ ±1% of reading or 0.02 NTU, whichever is greater
Resolution 0.01 / 0.1 / 1 NTU (FNU)
Stray light ≤ 0.02 NTU (FNU)
Calibration standard 0, 20, 100, 400, 800 NTU (FNU) AMCO Polymer or Formazin
Light detector Silicon photocell
Measuring modes Normal measurement and TruRead™ measurement
Data storage 200 measurement groups
Interface USB
Calibration memory Calibration date and time
Display TFT color display
Measuring cell ø25×60 mm, borosilicate glass with screw cap
Test volume 18 ml
Energy supply 3.7V rechargeable lithium battery
Operating conditions

Temperature: 0 - 50°C;rel. Humidity: 0 - 90% at 30°C,

0 - 80% at 40°C, 0 - 70% at 50°C, no condensation

Storage conditions

Gauge: -40 to 60°C (-40 to 140°F)

Calibration solutions: 5 to 30°C (41 to 86°F)

IP rating IP67
Security guidelines ISO9001:2015 and CE

Dimensions (WxLxH)

/Weight

Meter: (90×203×80) mm/ 385 g

Transport case: (310×295×110) mm/ 1.5 kg

 

A complete set contains:

  • TN500
  • Calibration solutions (0.00 NTU, 20.0 NTU, 100 NTU, 800 NTU, 800 NTU)
  • 6 sample cuvettes
  • Silicone oil
  • Microfiber cloth
  • USB stick with software
  • USB power supply
  • manual
  • Carrying case
The TN500 is CE certified and has a 2 year warranty.
Description value
WEEE obligation: Yes
WEEE registration number: DE 28154958
WEEE brand according to EAR register: APERA Instruments (Europe) GmbH
WEEE type of equipment: Small equipment, that can be used in private households

T500-1 Polymer AMCO Turbidity Calibration Solution Kit for TN500 and TN420 Standard Calibration...

Die T500-2 AMCO 0 NTU Standard-Kalibrierungslösung 100ml Nachfüllpackung 0 NTU Nachfülllösung...

T500-3 Replacement sample vials Replacement sample vials with lid for use with Apera...

T500-4 rechargeable lithium battery for TN500 and TN420 The rechargeable lithium battery for the...

T500-5 Tungsten Filament Lamp for Turbidity Meters TN500 and TN420 The replacement tungsten...